ZHAO Yan-hong, WANG Qiang, CHEN Zhao-mei. Determination of arginine in human fibrinogen by HPLC with precolumn derivatization[J]. Journal of Pharmaceutical Practice and Service, 2006, (5): 270-271.
Citation: ZHAO Yan-hong, WANG Qiang, CHEN Zhao-mei. Determination of arginine in human fibrinogen by HPLC with precolumn derivatization[J]. Journal of Pharmaceutical Practice and Service, 2006, (5): 270-271.

Determination of arginine in human fibrinogen by HPLC with precolumn derivatization

  • Received Date: 2005-12-16
  • Objective: To develop a method for determination of arginine in human fibrinogen. Methods: The sample was mixed with alanine as the internal standard, derived by 2, 4-dinitrofluorobenzene.The mobil phase was composed of 50% acetonitrile-0.05 mol/L acetate buffer(35∶65).The UV detection wavelength was 362nm. Results: There was a good linear relationship with the range of 0.16~0.06mg/mL(r=0.9998), average recovery of arginine was 102.4%, RSD was 1.4%. Conclusion: This method is accu-rate, simple, cheap, and can be used for determination of arginine in human fibrinogen.
  • 加载中
通讯作者: 陈斌, [email protected]
  • 1. 

    沈阳化工大学材料科学与工程学院 沈阳 110142

  1. 本站搜索
  2. 百度学术搜索
  3. 万方数据库搜索
  4. CNKI搜索

Article Metrics

Article views(2129) PDF downloads(205) Cited by()

Related
Proportional views

Determination of arginine in human fibrinogen by HPLC with precolumn derivatization

Abstract: Objective: To develop a method for determination of arginine in human fibrinogen. Methods: The sample was mixed with alanine as the internal standard, derived by 2, 4-dinitrofluorobenzene.The mobil phase was composed of 50% acetonitrile-0.05 mol/L acetate buffer(35∶65).The UV detection wavelength was 362nm. Results: There was a good linear relationship with the range of 0.16~0.06mg/mL(r=0.9998), average recovery of arginine was 102.4%, RSD was 1.4%. Conclusion: This method is accu-rate, simple, cheap, and can be used for determination of arginine in human fibrinogen.

ZHAO Yan-hong, WANG Qiang, CHEN Zhao-mei. Determination of arginine in human fibrinogen by HPLC with precolumn derivatization[J]. Journal of Pharmaceutical Practice and Service, 2006, (5): 270-271.
Citation: ZHAO Yan-hong, WANG Qiang, CHEN Zhao-mei. Determination of arginine in human fibrinogen by HPLC with precolumn derivatization[J]. Journal of Pharmaceutical Practice and Service, 2006, (5): 270-271.

Catalog

    /

    DownLoad:  Full-Size Img  PowerPoint
    Return
    Return